Enmet ISA-44-2OD Uživatelský manuál Strana 19

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 32
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 18
ISA-44-2-OD ENMET Corporation
15
7. Put the cal cup of the calibration fixture over the sensor. Let the gas flow over the sensor until the alarm triggers.
Use Table 6 for the approximate allowable time period until alarm:
Table 7: Approximate Alarm Time
Gas or Vapor Time to Alarm
20 ppm CO 5 - 7 minutes
50 ppm CO 5 minutes
200 ppm CO 2 minutes
10% LEL methane 1 minute
20% LEL methane 1 minute
100 ppm vinyl chloride 1 - 2 minutes
typical hydrocarbon vapors 1 - 2 minutes
20% LEL propane 1 minute
8. After you verify the alarm response, turn the gas off and remove the calibration fixture from the gas cylinder.
9. Turn the switch back to OPERATE.
CAUTION: If steps 1 - 10 do not trigger the alarms, and no electrical or mechanical malfunction is apparent, you must
recalibrate the instrument.
Zobrazit stránku 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 31 32

Komentáře k této Příručce

Žádné komentáře