Enmet EX-5175 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Enmet EX-5175. ENMET EX-5175 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 20
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
ENMET
Corporation
PO Box 979
Ann Arbor, MI 48106-0979
www.enmet.com
Manual Part Number
80003-099
MCN-13-008, 11/06/13
EX-5175-EC
Sensor/Transmitter
Manual
Zobrazit stránku 0
1 2 3 4 5 6 ... 19 20

Shrnutí obsahu

Strany 1 - Sensor/Transmitter

ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 www.enmet.com Manual Part Number 80003-099 MCN-13-008, 11/06/13

Strany 2

EX-5175-EC ENMET Corporation 6 4.0 Operation of the EX-5175-EC It is best to have the EX-5175-EC transmitters powered up and operational for 24 ho

Strany 3 - List of Figures and Tables

EX-5175-EC ENMET Corporation 7 5.0 Maintenance of the EX-5175-EC CAUTION: Do not open the EX-5175-EC S/T in a classified area. CAUTION: Do Not Att

Strany 4

EX-5175-EC ENMET Corporation 8 Regulator Gas Cylinder Select Menu Output, for Gas Input, for Calibration or Optional Gas Sampling Calibration Cov

Strany 5 - 1.0 Introduction

EX-5175-EC ENMET Corporation 9 5.2.1 Zero Adjust Zero function is Not applicable for Oxygen sensor/transmitters, go to Section 5.2.2 Gas Span. A Z

Strany 6

EX-5175-EC ENMET Corporation 10 Normal Display Mode 5ppm MENU HOLD the magnet over the MENU switch for 2 – 4 seconds to enter the Maintenan

Strany 7

EX-5175-EC ENMET Corporation 11 5.3 Sensor Replacement CAUTION: Area must be declassified during sensor replacement. Sensors should be replaced wh

Strany 8 - DC power

EX-5175-EC ENMET Corporation 12 6.0 Replacement Part Numbers ENMET replacement part numbers: Description Part Number For EX-5175-EC p/n Sensor: O

Strany 9 - S/W X.X

PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instru

Strany 11

Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.7

Strany 14 - Normal Display Mode

Table of Contents 1.0 INTRODUCTION ...

Strany 16 - 7.0 WARRANTY

EX-5175-EC ENMET Corporation 1 1.0 Introduction The ENMET EX-5175-ELECTROCHEMICAL, EC sensor/transmitters (S/T) is a 3-wire, 24 VDC 4-20mA S/T fo

Strany 17

EX-5175-EC ENMET Corporation 2 2.0 Features of the EX-5175-EC See Figure 1 for location of features: Feature Description Display LCD: Indicates th

Strany 18

EX-5175-EC ENMET Corporation 3 3.0 Installation of the EX-5175-EC CAUTION: Area must be declassified during installation. The ENMET EX-5175-EC gas

Strany 19 - Repair Return Form

EX-5175-EC ENMET Corporation 4 3.2 Wiring the EX-5175-EC to a Control Unit CAUTION: Area must be declassified during installation. Run conduit and

Strany 20

EX-5175-EC ENMET Corporation 5 3.3 Start up When the EX-5175-EC transmitter is first powered up, it goes through a series of momentary screens, wh

Komentáře k této Příručce

Žádné komentáře