3.7 Alarm Information
Alarm type Information displayed
Low alarm:
Slow Pulsed tone alarm sound
Alarm LED Flashing
Vibration
High alarm:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Vibration
STEL alert:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Vibration
TWA alert:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Vibration
Overrange alert:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Battery low voltage alert:
log appears.
At this time, the detector can still work for 30
days at least. When the battery is used up, the
detector will turn off automatically.
Sensor end-of-life indication:
Within 0-9 days before the sensor is used up,
screen will display as the right picture shows
when the detector is self testing after turning on.
The figure means the number of the days
remaining.
Table 1: RECON-IS Range and Alarm Levels
Target gas Detecting Range Low alarm level High alarm level TWA level STEL level
H
2
S 0-100ppm 10ppm 15ppm 10ppm 15ppm
CO 0-1000ppm 35ppm 100ppm 25ppm 200ppm
O
2
0-30%vol 19.5%vol 23.5%vol -- --
Zobrazit stránku 7