4.0 Display Settings
To display the current settings of the GS-24-DF use the push button switches. See figures 1 and 2 for
location of switches.
Access the M
ENU and SELECT switches through holes in front. See figures 1 and 2.
To cycle through the display settings press:
M
ENU switch once = Low Flow alarm setting.
M
ENU switch twice = High Flow alarm setting.
MENU switch three times = Audio alarm on or off.
M
ENU switch four times = Raw sensor signal in millivolts.
To disable audio alarm press:
S
ELECT switch once = Audio Alarm (horn) disabled for approximately 1 minute.
4.1 Changing Display Settings
To change the current setting of the GS-24-DF use the push button switches.
Press and hold M
ENU switch for three seconds, display will indicate Units
Press MENU switch again, once for each setting: SET LO, SET HI, Set audio or Display menu.
When Units is displayed: Press S
ELECT switch to change value.
Press M
ENU switch to step through the values: LPM, CCM, CFM
Press SELECT switch to exit, Display steps to SET LO
When SET LO is displayed: Press S
ELECT switch to display the value.
Press M
ENU switch to change the value.
Press SELECT switch to Exit
When SET HI is displayed: Press S
ELECT switch to display the value.
Press M
ENU switch to change the value.
Press SELECT switch to Exit
When SET AUD is displayed: Press S
ELECT switch to display On or Off.
Press MENU switch to change
Press SELECT switch to Exit
When DISPLAY is displayed: Press S
ELECT switch to display On or Off.
Press M
ENU switch for Flow Dsp or Go/NoGo.
Press SELECT switch to Exit