Gas Sampling Unit ENMET Corporation
7
5.0 Operation5.0 Operation
The Gas Sampling Unit requires no calibration itself, however any sensor/transmitter or instrument associated with it
does. Follow the instructions associated with the sensor/transmitter or instrument. To insure that the calibration is
accurate with the sample draw system, you need to fill a sample bag with calibration gas.
5.1 Procedure for Sampling from Gas Bag
Refer to Figures 6,7 and 8 for filling gas bag and attaching to Gas Sampling Unit.
Attach the regulator to the gas cylinder. Refer to Figure 6. Use only the regulator supplied with the calibration kit.
Attach a completely deflated gas bag (5 liter tedlar bag) to the output port of the regulator with appropriate tubing.
Turn on the gas cylinder by rotating the knob on the regulator so gas flows into the bag. Fill the gas bag until it is
approximately 75% inflated. Close the stopcock and turn off the regulator. Remove the gas cylinder regulator and
tubing.
Open the stopcock on the gas bag; connect the gas bag to the gas inlet port located on the Gas Sampler Unit, or the
entrance to the sensor adapter. Refer to Figures 7 and 8 for venting of calibration gas.
Allow gas to flow for the time required for calibration of the sensor/transmitter instrument.
Figure 6: Filling Sample Bag from Gas Cylinder
Sample Bag
Tubing
Regulator
Gas Cylinder
Stopcock